Table 5

Sensitivity of Attest rapid readout ethylene oxide biological indicator

Number
Sterilization
Incubation

growth-positives
False-negatives
Sensitivity
process temperature Number tested (168 hr) (4 hr) (4 hr)
37°C 600 mg
EO/L, 60% relative humidity 37°C 1,100 752 0 100%
54°C 600 mg
EO/L, 60% relative humidity 37°C 1,300 842 0