Figure 2

Figure 2. Enhanced view of receiver operating characteristic curves of 3 infrared thermal detection systems for detecting fever (oral temperature >100°F) showing sensitivities at false-positive rates (FPR) of 1%, 5%, and 10%. Red, FLIR ThermoVision A20M (FLIR Systems Inc., Boston, MA, USA); blue, OptoTherm Thermoscreen (OptoTherm Thermal Imaging Systems and Infrared Cameras Inc., Sewickley, PA, USA); and green, Wahl Fever Alert Imager HSI2000S (Wahl Instruments Inc., Asheville, NC, USA).