Volume 14, Number 8—August 2008
Dispatch
Cutaneous Infrared Thermometry for Detecting Febrile Patients
Table 1
Characteristic | Predicted tympanic temperature, °C† | ||
---|---|---|---|
>37.5 | >38.0 | >38.5 | |
Cutaneous temperature threshold, °C‡ | 37.5 | 38.0 | 38.5 |
Sensitivity | 0.76 (0.69–0.82) | 0.82 (0.71–0.90) | 0.82 (0.67–0.91) |
Specificity | 0.65 (0.63–0.67) | 0.77 (0.76–0.79) | 0.90 (0.88–0.91) |
Positive predictive value | 0.16 (0.14–0.19) | 0.10 (0.08–0.13) | 0.13 (0.09–0.18) |
Negative predictive value | 0.97 (0.96–0.98) | 0.99 (0.99–1.00) | 1.00 (0.99–1.00) |
Accuracy | 0.66 (0.64–0.68) | 0.78 (0.76–0.79) | 0.90 (0.89–0.91) |
*Values in parenthesis are 95% confidence intervals.
†Definition of hyperthermia.
‡Corresponds to the best threshold for a definition of cutaneous hyperthermia determined by using receiver operating characteristic curve.
Page created: July 12, 2010
Page updated: July 12, 2010
Page reviewed: July 12, 2010
The conclusions, findings, and opinions expressed by authors contributing to this journal do not necessarily reflect the official position of the U.S. Department of Health and Human Services, the Public Health Service, the Centers for Disease Control and Prevention, or the authors' affiliated institutions. Use of trade names is for identification only and does not imply endorsement by any of the groups named above.