Volume 31, Number 6—June 2025
Research
Diagnostic Accuracy of 3 Mpox Lateral Flow Assays for Antigen Detection, Democratic Republic of the Congo and United Kingdom
Suggested citation for this article
Abstract
The ongoing outbreaks of mpox highlight the urgent need for a rapid and low-cost diagnostic test to accurately detect and control this emerging disease. We estimated the analytical sensitivity using viral culture of the monkeypox virus clade IIb lineage and clinical diagnostic performance of 3 antigen detection rapid diagnostic tests (Ag-RDT) by using skin swab samples and upper-respiratory swab samples from mpox patients in the Democratic Republic of the Congo and the United Kingdom. The analytical limit of detection was 1.0 × 104 plaque-forming units/mL, fulfilling World Health Organization recommendations. Specificity of the 3 Ag-RDTs was 100%, but sensitivity was estimated at 0.00%–15.79% using skin samples and 0.00% using respiratory samples. None of the 3 Ag-RDTs reached the World Health Organization’s target clinical sensitivity, and we do not recommend them as diagnostic or screening tools for suspected mpox cases. Accurate Ag-RDTs for mpox diagnosis remain urgently needed.
Mpox is a zoonotic disease caused by monkeypox virus (MPXV), which belongs to the Orthopoxvirus genus, Poxviridae family; it has 2 major clades: clade I, which is subdivided into subclade Ia and Ib; and clade II, which is subdivided into subclade IIa and IIb (1). Historically, mpox was endemic to Central and West Africa. In May 2022, the number of mpox cases increased in a surge that included rapid expansion in nonendemic countries; it was declared the first mpox public health emergency of international concern (PHEIC) by the World Health Organization (WHO) (2). Since then, >100,000 cases of mpox and >200 deaths have been described in >120 countries not previously considered mpox-endemic. The number of infections during the 20th century has already been surpassed by the number of cases that occurred after the 2022 outbreak caused by clade II (3). On August 14, 2024, a second mpox PHEIC was declared by WHO after the substantial increase in mpox cases in the Democratic Republic of the Congo (DRC) and neighboring countries (4). In 2024, DRC, where mpox was first identified in 1970, reported the highest number of suspected cases globally, >27,000, and 800 deaths (5).
In the United Kingdom, most cases before 2022 were associated with travel from mpox-endemic countries. During August 2018–September 2021, a total of 7 mpox cases were identified in the United Kingdom (4 imported cases and 3 secondary cases) (6). The discovery of the first mpox case of the 2022 global outbreak in the United Kingdom was on May 7, 2022, in a person who traveled from Nigeria; as of June 8, 2022, a total of 336 cases had been laboratory confirmed, a number that rose to 3,732 by the end of 2022. On October 30, 2024, the first clade Ib mpox case was confirmed in the United Kingdom; 3 further household contact cases were confirmed (7). Those were the first locally transmitted clade Ib mpox cases in the WHO European Region and the first outside Africa since a PHEIC was declared for a second time in August 2024 (4).
To confirm a clinical diagnosis, WHO advises testing for MPXV as soon as possible in persons who fit the suspected case definition. Laboratory-based real-time PCR is the primary method used for MPXV detection. Laboratory-based PCR testing requires costly equipment, up-front DNA extraction, and skilled personnel, which might only be available in specialized laboratories, making rapid detection of cases during outbreaks more challenging. In contrast, rapid diagnostic tests (RDTs) are low cost, equipment free, easy to use, and suitable to use at the point of care (POC); results are available within 20 minutes (8). The value of antigen-detecting RDTs (Ag-RDTs) in rapidly detecting infected persons and enabling isolation and management of patients has been proven for many viral diseases, notably during the COVID-19 pandemic (9).
The global increase in mpox cases after the 2022 PHEIC and the subsequent PHEIC 15 months later brought to light the increased demand for decentralized POC diagnostics for this highly infectious virus and highlighted the urgent need for Ag-RDTs for MPXV as a priority. This increased demand has resulted in the availability of dozens of Ag-RDTs in the market. As of January 2024, >69 Ag-RDTs for MPXV were in varying stages of development, of which >35 had received CE marking for in vitro diagnostics (IVD) (required for devices sold in Europe) and were commercially available (10). Despite the increased number of Ag-RDTs for MPXV, clinical evaluation data are still lacking (11). To ensure reliable and accurate performance of Ag-RDTs, diagnostic evaluation studies across multiple, independent sites are required to generate evidence of their effectiveness to guide implementation.
The aim of this study was to evaluate the diagnostic performance of 3 Ag-RDT brands at detecting MPXV antigens: FlowFlex Monkeypox Virus Antigen Rapid Test (ACON Biotech [Hangzhou] Co., Ltd., https://www.aconbio.com), Ecotest Monkeypox Antigen Rapid Test (Assure Tech [Hangzhou] Co., Ltd., https://www.assuretech-product.com), and Standard Q Monkeypox Ag Test (SD Biosensor, https://www.sdbiosensor.com). We used skin lesion swab samples and upper respiratory tract swab samples from patients in DRC and the United Kingdom, 2 countries with different MPXV epidemiologic characteristics and clades (clade I [DRC] and clade II [United Kingdom]). We evaluated the Ag-RDTs in prospectively collected samples in DRC and retrospectively in the United Kingdom.
Study Settings and Participants
In DRC, persons >2 years of age suspected to have mpox according to the WHO case definition (12) were eligible to participate in the study. Ethical approval was obtained by DRC’s National Ethics and Health Committee (Comité National d’Ethique et de la Santé [CNES], reference 452/CNES/BN/PMMF/2023). Recruitment took place during January–December 2023 in Maniema Province through home visits of eligible persons. Paired skin samples and upper respiratory specimens were collected from all recruited participants (n = 68) by trained heathcare workers and placed in 3 mL of noninactivating virus transport medium (VTM) for viral preservation. The Ag-RDTs could not be tested at the POC because of health and safety restrictions; all VTM samples were transported in cool boxes (2°C–8°C) to the Institut National de Recherche Biomédicale (INRB) Biosafety Level 2 laboratories in Lubutu for processing according to national guidance for MPXV testing. All VTM tubes were processed within 4 hours for MPXV Ag-RDT testing and quantitative PCR (qPCR).
In the United Kingdom, we used retrospectively collected skin samples (n = 30) and upper respiratory samples (n = 23 [1 nasopharyngeal, 22 oropharyngeal]) in universal transport media (UTM) (Copan, https://www.copangroup.com) from a cohort of 16 adult patients positive for mpox by PCR from the Royal Liverpool University Hospital, Sheffield Teaching Hospital NHS Foundation Trust, and Royal Free London Hospital for this study. Patients were recruited during the last 2 outbreaks of mpox in the United Kingdom in 2018 and 2022. Trained healthcare workers collected all swab samples. Patients gave consent under the WHO ISARIC Clinical Characterization Collaboration Protocol for severe emerging infections (ISRCTN66726260). Ethical approval was obtained from the National Research Ethics Service and the Health Research Authority (IRAS ID:126600, REC 13/SC/0149).
Mpox diagnosis was confirmed by the UK Health Security Agency using qPCR before enrollment in the study. In addition to samples from mpox-positive patients, to fulfill the minimum number of negative swab specimens for mpox diagnostic evaluations recommended by the US Food and Drug Administration (13), we used a set of 32 leftover nasopharyngeal samples in UTM from previous COVID-19 studies (14) as mpox negative controls. The samples were collected under the Facilitating Accelerated Clinical Validation of Novel diagnostics for COVID-19 (FALCON), and ethical approval was obtained from the National Research Ethics Service and the Health Research Authority (IRAS ID:28422, REC: 121 20/WA/0169). All samples were aliquots stored at –80°C and thawed for the first time for this study. Samples were processed and tested at the Biosafety Level 3 laboratories of the Liverpool School of Tropical Medicine (LSTM) as previously described (14).
MPXV Ag-RDT testing
We selected the Ag-RDTs evaluated in this study after an expression of interest launched by FIND (https://www.finddx.org) and a scoring process based on defined criteria. We evaluated 3 Ag-RDTs: FlowFlex Monkeypox Virus Antigen Rapid Test, Ecotest Monkeypox Antigen Rapid Test, and Standard Q Monkeypox Ag Test. The 3 RDTs are based on immunochromatography and show the presence of MPXV A29L antigen using colloidal gold for visualization. Flowflex and Ecotest were commercially available, whereas Standard Q was for research use only at the time of evaluation. All test brands can be used with skin lesion samples. In addition, Flowflex can be used with serum, plasma, and upper respiratory samples; Standard Q can be used in serum, plasma, and whole-blood samples; and Ecotest can be used in upper respiratory samples.
We performed tests in INRB Biosafety Level 2 laboratories in DRC and in LSTM Biosafety Level 3 laboratories in the United Kingdom. In brief, we added the specified amount of VTM or UTM confirmed by the manufacturers (200 μL for Flowflex and Ecotest and 300 μL for Standard Q) into the extraction buffer and then added the number of drops of the extraction buffer specified in the instructions for use into the sample well (4 drops for Flowflex and Standard Q and 3 drops for Ecotest). We read tests and interpreted them visually after 15–30 minutes according to the instructions. Two independent technicians read the results; a third technician acted as a tiebreaker in case of discrepant results.
Reference MPXV qPCR Test
At both sites, we extracted DNA and performed MPXV qPCR using the same UTM or VTM tube used for the 3 Ag-RDT tests. At INRB, we extracted DNA from a 300 μL aliquot of sample by using the Natch 16S automated platform with the Nucleic Acid Extraction-Purification Kit (both Sansure Biotech, https://www.sansureglobal.com), according to the instructions for use. At LSTM, we extracted DNA from 200 μL of UTM using the QiAamp96 Virus Qiacube HT kit (QIAGEN, https://www.qiagen.com), according to the instructions for use.
We used the same MPXV qPCR reference test in both sites for evaluating index tests (Monkeypox Virus Nucleic Acid Diagnostic Kit; Sansure Biotech). We carried out qPCR by using a MA-1620Q qPCR thermocycler (Sansure Biotech) at INRB and a QuantStudio 5 (Thermo Fisher Scientific, https://www.thermofisher.com) at LSTM. We considered a qPCR result with a cycle threshold (Ct) <40 MPXV positive according to instructions for use. We used this qPCR kit as the reference test because it has been successfully demonstrated to detect MPXV clades I, IIa, and IIb (15), is CE marked for commercial use, and has demonstrated higher diagnostic accuracy than the mpox Centers for Disease Control and Prevention laboratory–based qPCR (16).
Analytical Limit of Detection of Ag-RDTs
We cultured mpox viral stock of a MPXV strain from clade II, subclade IIb, lineage B 0.1 (Slovenia_MPXV–1_2022) obtained from the European Virus Archive Global (https://www.european-virus-archive.com) in Vero E6 cells (ECACC 85020206) in Dulbecco’s Modified Eagle Medium plus 10% fetal bovine serum and 1% penicillin/streptomycin solution to generate the MPXV stock. We serially diluted a fresh aliquot 10-fold using UTM to produce concentrations from 5.0 × 104 to 5.0 × 100 PFU/mL. We defined the limit of detection (LOD) as the lowest concentration at which all 3 replicates were positive. Once the LOD was achieved, half dilutions were tested above and below the LOD. We performed Ag-RDT testing to calculate the LOD and quantified the viral copy numbers per mL (copies/mL) of the serial dilutions, as previously described (14,16).
Statistical Analysis
To assess the diagnostic accuracy of Ag-RDTs in patients with suspected mpox, we calculated point estimates of sensitivity and specificity for each Ag-RDT on the basis of results of the reference MPXV qPCR assay from the same VTM or UTM tube used for the Ag-RDT. We derived the 95% CI for each point estimate on the basis of Wilson’s score method. To compare performance of the Ag-RDTs at different Ct values, we stratified point estimates of sensitivity by Ct value of the reference test. We used 2-tailed Fisher exact test and χ2 test to determine nonrandom associations between categorical variables. We assessed differences between the Ct values (expressed as mean +SD) in sample groups using the paired Student t test. Statistical significance was set at <0.05. We performed statistical analysis using R version 4.5.0 (R Foundation for Statistical Computing, https://ww.r-project.org) and GraphPad Prism version 9.1.0 (GraphPad Software, Inc., https://www.graphpad.com).
Clinical Evaluation
In DRC, 34/68 (50%) of mpox patients were men (Table 1). The median time from onset of symptoms was 4 (range 1–34) days. The most prevalent symptoms were fever (91%), skin lesions (100%), influenza-like symptoms (75%), headaches (54%), and cough (50%). In the United Kingdom, 16/16 (100%) mpox patients were men; mean age was 35.1 (range 24–58) years. The median time from onset of symptoms was 8 (range 0–11) days. The most common symptoms were skin lesions (100%), skin rashes (87.5%), and fever (68.8%).
In DRC, 19/68 (28%) skin samples and 14/68 (21%) upper-respiratory samples from persons suspected of having mpox tested positive using the Sansure qPCR. Flowflex and Ecotest Ag-RDTs detected MPXV antigens in 3/19 MPXV-positive skin samples, resulting in a clinical sensitivity of 15.79% (95% CI 5.52%–37.57%), whereas Standard Q detected MPXV antigens in 2/19 samples, resulting in a clinical sensitivity of 10.53% (95% CI 2.94%–31.39%). The Ag-RDT Flowflex was more sensitive when detecting MPXV antigen in skin samples with Ct <20 than those with Ct values >33 (p = 0.008); however, this difference was not observed with the other Ag-RDT brands. None of the Ag-RDT brands detected MPXV antigen in upper respiratory samples, resulting in 0% (95% CI 0%–23.2%) sensitivity. The clinical specificity was 100% (95% CI 92.73%–100%) for each of the Ag-RDTs in both sample types (Table 2).
In the United Kingdom, 16/23 upper-respiratory samples (69.56%) and 27/30 skin samples (90%) from mpox-positive patients were positive by Sansure qPCR. All 32 upper respiratory samples analyzed from the COVID-19 cohort tested negative for MPXV as expected. No positive results were obtained when testing either respiratory or skin swab samples regardless of the Ag-RDT brand used 0% (95% CI 0%–20.59%). The specificity was 100% (95% CI 90.97%–100%) for the 3 Ag-RDT brands on both sample types (Table 3).
The difference in sensitivity in MPXV Ag-RDTs was lower when testing upper-respiratory samples than in skin samples (p = 0.007). We assessed the comparison of the Ct values and noted a difference in Ct values between upper-respiratory and skin sample groups (p=0.042) from DRC but not from the United Kingdom (Figure 1). The mean Ct value of upper-respiratory samples in DRC was 30.7 (+4.79) and mean Ct value for skin samples was 26.63 (+6.87), whereas in the United Kingdom mean Ct value for respiratory samples was 27.2 (+2.34) and for skin was 28.83 (+6.88). We found no difference in sensitivity between the 3 Ag-RDT brands and between countries. We also analyzed test results by onset of symptoms (Figure 2) but observed no difference in Ag-RDT results by symptom onset group.
Analytical Evaluation
Using the viral stock, all Ag-RDT brands were positive at 5.0 × 104 PFU/mL, 2.5 × 104 PFU/mL and 1.0 × 104 PFU/mL. The LOD of all the Ag-RDT brands using the MPXV viral culture was determined to be 1.0 × 104 PFU/mL (1.3 × 105 copies/mL). All concentrations tested below the LOD were negative in all instances.
After the recent PHEIC, WHO issued an urgent call to accelerate availability of POC diagnostics for mpox (17). The lack of validation data for MPXV Ag-RDTs represents a large gap in the diagnostics landscape that has slowed down rapid, effective responses to new outbreaks and ongoing endemic transmission (18). The primary aim of this study was to evaluate the diagnostic accuracy of 3 Ag-RDT brands (Flowflex, Ecotest, and Standard Q) in DRC and the United Kingdom.
WHO’s target product profile (TPP) for MPXV Ag-RDTs recommends minimal clinical sensitivity of 80% and specificity of 97% (19). Specificity was fulfilled by the 3 Ag-RDTs evaluated in both countries, but sensitivity was extremely low (0%–15.79%), making the tests unsuitable for diagnostic or screening use. Evaluation data on Ag-RDT for MPXV are very limited. A previous study reported detection of MPXV antigens using the Ag-RDT Tetracore Orthopox BioThreat (https://tetracore.com) in 5 of 6 tested MPXV-positive samples with low Ct values (Ct 15–22) (20). In addition to the limited number of samples, this assay required sonication for swab material and dry ice/ethanol bath freezing followed by pestle grinding, making it unsuitable for POC use. Another study using an Orthopoxvirus Ag-RDT prototype failed to detect MPXV antigens among 80 MPXV qPCR-positive clinical samples in Belgium (21). That study suggested that the failure to detect MPXV antigen in swab samples could be caused by inhibition by the inactivating components of the VTM, which can cause protein denaturation. In this study, we used noninactivating swab transport medium in both sites and different types of medium (VTM in the DRC and UTM in the United Kingdom); clinical sensitivity was not improved. Studies on Ag-RDTs for SARS-CoV-2 comparing the use of dry swabs with Amies, VTM, and UTM have documented false-positive results because of nonspecific electrostatic interactions between the antibodies in the assay (22,23), a decreased LOD because of a dilution effect (23), or no changes in sensitivity or specificity depending on the Ag-RDT brand (23). The use of different types of swab transport medium should be investigated to optimize performance of Ag-RDT for mpox while preserving the virus for transport and storage.
Investigations of the analytical sensitivity of these Ag-RDTs gave an LOD of 1.0×104 pfu/mL, being more sensitive than previous analytical evaluations of Ag-RDTs for MPXV. The Orthopoxvirus Ag-RDT prototype had an LOD of 3.0 × 105 PFU/mL (21), and the commercially available Tetracore Orthopox BioThreat had an LOD of 1.5 × 106 PFU/mL after sonication (20). The recommended analytical LOD in the WHO TPP is at 106 PFU/mL, being fulfilled by the 3 brands of Ag-RDTs evaluated here and the previously published study on the Orthopoxvirus Ag-RDT prototype (21), suggesting that laboratory sensitivity using the PFU/mL measurement does not align with clinical sensitivity in the field.
The use of LOD using viral isolates is often used as a proxy before having the test evaluated using clinical specimens; however, in this study and others (21), the correlation between analytical and clinical sensitivity for MPXV has been shown to be very poor, yielding lower sensitivity than expected among clinical samples. The reason for this variability in antigen detection sensitivity between mpox clinical samples and mpox viral isolates is still uncertain and needs further investigation, as does the quantity and type of accessible antigen in clinical samples. In addition, the targeted antigen of the Ag-RDTs evaluated in this study was MPXV A29L. Target antigens for other Ag-RDT brands include A29L, A35R, A5L, B6R, E8L, H3, and M1R (10). The antigen A27L (homologous of MPXV A29L in vaccinia virus) has previously been suggested to be a good candidate because it is conserved and abundant within the virion; however, the Ag-RDTS targeting this antigen in this study and reported elsewhere (21) did not yield acceptable sensitivity. This finding highlights the need for further evaluations using clinical samples with Ag-RDTs that target different antigen types. Currently, manufacturers of >3 Ag-RDT brands in the market have disclosed the use of MPXV A29L as antigen target (Hangzhou Testsea Biotechnology, Guangdong Wesail Biotech, and Nanjing Synthgene Medical Technology).
Flowflex and Ecotest are designed to be used with upper-respiratory samples; however, none of the Ag-RDTs detected MPXV antigens in those samples, suggesting that using this sample type for antigen detection is not appropriate. Diagnostic evaluation studies using PCR found lower positivity rates in respiratory samples than in skin samples (24), which might be attributed to lower viral titer levels (25) or earlier clearance in this sample type (26), which is exacerbated by the lower sensitivity of the Ag-RDTs evaluated here.
Major limitations of this study were that testing could not be done at the POC in DRC (because samples had to be transported to the designated laboratory) and that retrospective frozen samples were used in the United Kingdom. The effect of testing delay and of sample storage and freeze-thawing on Ag-RDT results has not been studied with MPXV. Studies on SARS-CoV-2 Ag-RDTs noted a decline in test-line intensity (not false-negative results) after storage periods of 24 hours to >7 days at 2°C–8°C (27–30). Results of Ag-RDT testing for SARS-CoV-2 did not find a significant difference between 101 datasets that involved fresh specimens and 23 freeze-thawed specimens (31). However, MPXV is a larger DNA virus, whereas SARS-CoV-2 is a smaller RNA virus; thus, the 2 are not directly comparable. The use of retrospective frozen and refrigerated samples is accepted for production of clinical diagnostic data (13), and the WHO TPP recommends that MPXV Ag-RDTs be compatible with samples that have been refrigerated or frozen with use of preservation media for quality control, repeats, or follow-up testing (19).
In conclusion, the results of this study raise considerable doubts on the suitability of Ag-RDT for mpox surveillance and diagnosis because of their poor clinical sensitivity among suspected mpox cases. Recommendations for future mpox Ag-RDT evaluations should include brands that detect different MPXV antigens and evaluation of different swab preservation mediums.
This article was published as a preprint at https://www.medrxiv.org/content/10.1101/2024.11.07.24316894v1.
Dr. Ishara-Nshombo is a scientist and project manager at the Institut National de Recherche Biomédicale in the Democratic Republic of the Congo. His work focuses on epidemiological surveillance of emerging diseases at the human–animal interface. Dr. Somasundaran works in diagnostic research at Liverpool School of Tropical Medicine Centre for Drugs and Diagnostics in the United Kingdom. Her primary research interests are public health and improving disease detection.
Acknowledgments
We thank the FIND team dedicated to managing and supervising this study: Emmanuel Agogo, Audrey Albertini, Daniel G. Bausch, Berra Erkosar, Susan Logoose, Michael Otieno, and Aurelia Vessiere. In the Democratic Republic of the Congo, we thank all the participants who volunteered for this study. We also thank all the staff at Lubutu General Referral Hospital, especially Jean Meniko, Fiston Kaleba, Berthon Okoma, Naomie Koho, Emmanuel Matengano, and Justin Kapela, for their contribution to the success of this study. We would also like to thank Patrick Karungu for his work in entering the source data into the electronic database. In the United Kingdom, we thank the ISARIC 4C investigators (https://isaric4c.net/about/authors), all the Clinical Research Network research nurses, and clinical research fellows for supporting us with the sample collection and recruitment. We thank the CONDOR steering group: A. Joy Allen, Julian Braybrook, Richard Body, Peter Buckle, Eloise Cook, Paul Dark, Kerrie Davis, Gail Hayward, Adam Gordon, Anna Halstead, Charlotte Harden, Colette Inkson, Naoko Jones, William Jones, Dan Lasserson, Joseph Lee, Clare Lendrem, Andrew Lewington, Mary Logan, Massimo Micocci, Brian Nicholson, Rafael Perera-Salazar, Graham Prestwich, D. Ashley Price, Charles Reynard, Beverley Riley, John Simpson, Valerie Tate, Philip Turner, Mark Wilcox, and Melody Zhifang.
The raw data for this evaluation are available from the corresponding authors upon reasonable request.
This work was funded as part of FIND’s work as coconvener of the diagnostics pillar of the Pandemic Threats Programme, who received funds from the Germany (Federal Ministry of Economic Cooperation and Development) and Canada governments. ISARIC4C was funded from the National Institute for Health Research (NIHR) (award CO-CIN-01), the UK Medical Research Council (grant MC_PC_19059), the NIHR for Health Research Health Protection Research Unit (HPRU) in Emerging and Zoonotic Infections (in partnership with the University of Liverpool), the United Kingdom Health Security Agency (UKHSA), LSTM and the University of Oxford) (NIHR award 200907), the Wellcome Trust and Department for International Development (215091/Z/18/Z), Bill and Melinda Gates Foundation (OPP1209135), and the Liverpool Experimental Cancer Medicine Centre (grant reference C18616/A25153). The FALCON study was funded by the NIHR, Asthma United Kingdom, and the British Lung Foundation.
The study was planned and designed by A.I.C.-A., J.D., D.M.E., H.K.-M., E.I.-N., D.M.E., and D.M.-B. Laboratory work was performed by A.S., J.P., Y.H., D.W., K.K., C.T.W., A.R.-R., and E.M.-M. under the supervision of A.I.C.-A., T.E., E.I.-N., and H.K.-M. Data analysis and interpretation were conducted by A.S., A.R.-R., A.I.C.-A., M.W., and E.I.-N. A.I.C.-A. and E.I.-N. drafted the initial manuscript. The funding was obtained by A.I.C.-A., H.K.-M., M.G.S., J.K.B., and J.D. Study oversight was provided by A.I.C.-A. and H.K.-M. All authors reviewed and approved the final manuscript.
Members of ISARIC 4C: Mike Beadsworth, Ingeborg Welters, Lance Turtle, Jane Minton, Karl Ward, Elinor Moore, Elaine Hardy, Mark Nelson, David Brealey, Ashley Price, Brian Angus, Graham Cooke, and Oliver Koch.
References
- World Health Organization. Mpox. 2024 [cited 2025 Mar 16]. https://www.who.int/news-room/fact-sheets/detail/mpox
- Sah R, Padhi BK, Siddiq A, Abdelaal A, Reda A, Ismail Lashin B, et al. Public health emergency of international concern declared by the World Health Organization for monkeypox. Glob Secur Health Sci Policy. 2022;7:51–6. DOIGoogle Scholar
- World Health Organization. Global mpox trends [cited 2024 Oct 15]. https://worldhealthorg.shinyapps.io/mpx_global
- World Health Organization. WHO Director-General declares mpox outbreak a public health emergency of international concern [cited 2024 Oct 9]. https://www.who.int/news/item/14-08-2024-who-director-general-declares-mpox-outbreak-a-public-health-emergency-of-international-concern
- Breman JG, Kalisa-Ruti , Steniowski MV, Zanotto E, Gromyko AI, Arita I. Human monkeypox, 1970-79. Bull World Health Organ. 1980;58:165–82.PubMedGoogle Scholar
- Gov.UK. Mpox outbreak: epidemiological overview, 6 February 2025 [cited 2025 Mar 16]. https://www.gov.uk/government/publications/monkeypox-outbreak-epidemiological-overview/mpox-outbreak-epidemiological-overview-6-february-2025
- Gov.UK. Confirmed cases of mpox clade Ib in United Kingdom [cited 2025 Mar 16]. https://www.gov.uk/guidance/confirmed-cases-of-mpox-clade-ib-in-united-kingdom
- Land KJ, Boeras DI, Chen XS, Ramsay AR, Peeling RW. REASSURED diagnostics to inform disease control strategies, strengthen health systems and improve patient outcomes. Nat Microbiol. 2019;4:46–54. DOIPubMedGoogle Scholar
- Drain PK. Rapid diagnostic testing for SARS-CoV-2. N Engl J Med. 2022;386:264–72. DOIPubMedGoogle Scholar
- FIND. Monkeypox diagnostic test directory now live [cited 2024 Oct 14]. https://www.finddx.org/publications-and-statements/monkeypox-diagnostic-test-directory-now-live/
- Lim CK, Roberts J, Moso M, Liew KC, Taouk ML, Williams E, et al. Mpox diagnostics: Review of current and emerging technologies. J Med Virol. 2023;95:
e28429 . DOIPubMedGoogle Scholar - Adler H, Gould S, Hine P, Snell LB, Wong W, Houlihan CF, et al.; NHS England High Consequence Infectious Diseases (Airborne) Network. Clinical features and management of human monkeypox: a retrospective observational study in the UK. Lancet Infect Dis. 2022;22:1153–62. DOIPubMedGoogle Scholar
- US Food and Drug Administration. Template for developers of antigen diagnostic tests for monkeypox. Emergency use authorizations for medical devices [cited 2025 Jan 2]. https://www.fda.gov/medical-devices/emergency-use-authorizations-medical-devices/monkeypox-mpox-emergency-use-authorizations-medical-devices#templates
- Byrne RL, Aljayyoussi G, Kontogianni K, Clerkin K, McIntyre M, Wardale J, et al. Head-to-head comparison of anterior nares and nasopharyngeal swabs for SARS-CoV-2 antigen detection in a community drive-through test centre in the UK. BMJ Open Respir Res. 2025;12:
e001747 . DOIPubMedGoogle Scholar - Bunse T, Ziel A, Hagen P, Rigopoulos G, Yasar U, Inan H, et al. Analytical and clinical evaluation of a novel real-time PCR-based detection kit for Mpox virus. Med Microbiol Immunol. 2024;213:18. DOIPubMedGoogle Scholar
- Romero-Ramirez A, Somasundaran A, Kontogianni K, Parkes J, Hussain Y, Gould S, et al.; CONDOR steering group; ISARIC 4C Investigators. Evaluation of the diagnostic accuracy of Xpert® Mpox and STANDARD™ M10 MPX/OPX for the detection of monkeypox virus. J Infect. 2025;90:
106413 . DOIPubMedGoogle Scholar - International Pandemic Preparedness Secretariat. Mpox day 15: the status of rapid diagnostic tests, Tecovirimat trials and Emergency Use Listings [cited 2024 Oct 9]. https://ippsecretariat.org/news/mpox-day-15
- Unitaid. Mpox emergency response must prioritize access to diagnostics, clinical care and vaccines [cited 2024 Oct 9]. https://unitaid.org/news-blog/mpox-emergency-response-must-prioritize-access-to-diagnostics-therapeutics-and-vaccines/#en
- World Health Organization. Target product profiles for tests used for mpox (monkeypox) diagnosis [cited 2025 Jan 4]. https://www.who.int/publications/i/item/9789240076464
- Townsend MB, MacNeil A, Reynolds MG, Hughes CM, Olson VA, Damon IK, et al. Evaluation of the Tetracore Orthopox BioThreat® antigen detection assay using laboratory grown orthopoxviruses and rash illness clinical specimens. J Virol Methods. 2013;187:37–42. DOIPubMedGoogle Scholar
- Laidlaw SM, Ulaeto D, Lonsdale S, Clark G, Sumner R, Edwards T, et al. Detection of mpox and other orthopoxviruses using a lateral flow device as a point-of-care diagnostic. Microbiol Spectr. 2025;13:
e0245624 . DOIPubMedGoogle Scholar - Mayfield J, Hesse P, Ledden D. The impact of universal transport media and viral transport media liquid samples on a SARS-CoV-2 rapid antigen test. Arch Intern Med Res. 2022;5:481–7. DOIGoogle Scholar
- Cubas-Atienzar AI, Kontogianni K, Edwards T, Wooding D, Buist K, Thompson CR, et al. Limit of detection in different matrices of 19 commercially available rapid antigen tests for the detection of SARS-CoV-2. Sci Rep. 2021;11:18313. DOIPubMedGoogle Scholar
- Zhou YB. Skin lesion specimens as first choice to detect monkeypox virus. Lancet. 2023;401:1264–5. DOIPubMedGoogle Scholar
- Tarín-Vicente EJ, Alemany A, Agud-Dios M, Ubals M, Suñer C, Antón A, et al. Clinical presentation and virological assessment of confirmed human monkeypox virus cases in Spain: a prospective observational cohort study. Lancet. 2022;400:661–9. DOIPubMedGoogle Scholar
- Suñer C, Ubals M, Tarín-Vicente EJ, Mendoza A, Alemany A, Hernández-Rodríguez Á, et al.; Movie Group. Viral dynamics in patients with monkeypox infection: a prospective cohort study in Spain. Lancet Infect Dis. 2023;23:445–53. DOIPubMedGoogle Scholar
- Mertens P, De Vos N, Martiny D, Jassoy C, Mirazimi A, Cuypers L, et al.; LHUB-ULB SARS-CoV-2 Working Diagnostic Group. Development and potential usefulness of the COVID-19 Ag Respi-Strip Diagnostic Assay in a pandemic context. Front Med (Lausanne). 2020;7:225. DOIPubMedGoogle Scholar
- Puyskens A, Bayram F, Sesver A, Michel J, Krause E, Bourquain D, et al. Performance of 20 rapid antigen detection tests to detect SARS-CoV-2 B.1.617.2 (Delta) and B.1.1.529 (Omicron) variants using a clinical specimen panel from January 2022, Berlin, Germany. Euro Surveill. 2023;28:
2200615 . DOIPubMedGoogle Scholar - Gokulan CG, Kiran U, Kuncha SK, Mishra RK. Temporal stability and detection sensitivity of the dry swab-based diagnosis of SARS-CoV-2. J Biosci. 2021;46:1–9. DOIPubMedGoogle Scholar
- Milo R, Phillips R. How fast do RNAs and proteins degrade? [cited 2025 Jan 3]. https://book.bionumbers.org/how-fast-do-rnas-and-proteins-degrade
- Parvu V, Gary DS, Mann J, Lin YC, Mills D, Cooper L, et al. Factors that influence the reported sensitivity of rapid antigen testing for SARS-CoV-2. Front Microbiol. 2021;12:
714242 . DOIPubMedGoogle Scholar
Figures
Tables
Suggested citation for this article: Ishara-Nshombo E, Somasundaran A, Romero-Ramirez A, Kontogianni K, Mukadi-Bamuleka D, Mukoka-Ntumba M, et al. Diagnostic accuracy of 3 mpox lateral flow assays for antigen detection, Democratic Republic of the Congo and United Kingdom. Emerg Infect Dis. 2025 Jun [date cited]. https://doi.org/10.3201/eid3106.250166
Original Publication Date: May 02, 2025
1These authors contributed equally to this article.
2Members of ISARIC 4C investigator group are listed at the end of this article.
Table of Contents – Volume 31, Number 6—June 2025
EID Search Options |
---|
|
|
|
Please use the form below to submit correspondence to the authors or contact them at the following address:
Ana Isabel Cubas-Atienzar, Centre for Drugs and Diagnostics, Liverpool School of Tropical Medicine, Pembroke Place, Liverpool L3 5QA, UK
Top