Volume 7, Number 2—April 2001
THEME ISSUE
4th Decennial International Conference on Nosocomial and Healthcare-Associated Infections
State of the Art

New Disinfection and Sterilization Methods

William A. RutalaComments to Author  and David J. Weber
Author affiliations: University of North Carolina (UNC) Health Care System; UNC School of Medicine, Chapel Hill, North Carolina, USA

Main Article

Table 5

Sensitivity of Attest rapid readout ethylene oxide biological indicator

Number
Sterilization
Incubation

growth-positives
False-negatives
Sensitivity
process temperature Number tested (168 hr) (4 hr) (4 hr)
37°C 600 mg
EO/L, 60% relative humidity 37°C 1,100 752 0 100%
54°C 600 mg
EO/L, 60% relative humidity 37°C 1,300 842 0

Main Article

Page created: April 17, 2012
Page updated: April 17, 2012
Page reviewed: April 17, 2012
The conclusions, findings, and opinions expressed by authors contributing to this journal do not necessarily reflect the official position of the U.S. Department of Health and Human Services, the Public Health Service, the Centers for Disease Control and Prevention, or the authors' affiliated institutions. Use of trade names is for identification only and does not imply endorsement by any of the groups named above.